TBS-380 Fluorometer
Minicell Adaptor Kit

  • Measure sample volumes as low as 50 µL without losing sensitivity.
  • Adaptor fits into the TBS-380 Mini-Fluorometer sample compartment.
  • Easy to use.

Overview

For applications where sample volume is a concern, the Minicell Adaptor with two focusing lenses allows you to measure sample in an assay volumes as low as 50 µL without losing sensitivity (e.g. 1 ng/mL dsDNA using PicoGreen, 50 pg of dsDNA in 50 µL assay volume).

The Minicell Adaptor Kit comes equipped with 400 borosilicate, flat-bottomed vials that are easily loaded in and out of the adaptor. The disposable vials eliminate potential cross-contamination and time-consuming cleanup required with non-disposable cells. Vials accommodate a sample volume range of 50-250 µL.


Results from the Minicell Adaptor used to quantitate DNA with Molecular Probes PicoGreen Assay.

Minimum assay volume
10 x 10 mm cuvette
2 mL
Minicell
50 µL
Minimum Detection Limit 500 pg/mL
1 ng/mL
Minimum amount of DNA required 1 ng 50 pg
Cost of the reagent per sample $1.20 $0.03

Table 1. Benefit of using Minicell with the PicoGreen Assay.

 