Modulus II Multimode Microplate Reader A modular microplate multimode reader for luminescence, fluorescence, and absorbance Life Science applications. Overview The Modulus™ II Multimode Microplate Reader is skillfully designed for today’s life science laboratory. In addition to having performance on par with single-mode instruments, the Modulus™ II Microplate blends user-friendly operation with easy data handling and flexible purchasing options. The result of this design is an instrument with superior performance that is easy to use, is affordable, and can be customized to your laboratory’s needs. Performance The Modulus™ II Multimode Microplate Reader combines the superior performance expected from single-mode instruments with the functionality of multiple modes. To achieve industry-leading performance, the Modulus™ II Microplate is designed with optical channels dedicated to each individual technology. Unlike other multimode systems, readings taken with the Modulus™ II Microplate are not degraded by indirect fiber-optic transmission or crowded optical channels. Dedicated optical channeling ensures that the Modulus™ II Microplate provides sensitivity and dynamic range on par with the highest performing single-mode instruments. Ease of Use The Modulus™ II Multimode Microplate Reader is designed to be put into use straight from the box, without the need to read a manual or obtain special training. To achieve this plug-and-play usability, the Modulus™ II Microplate combines a color touch screen display with a Windows-based onboard computer. The built-in computer eliminates the connectivity hassles, cost, and space requirements of running an external computer. Assay set-up on the Modulus™ II Microplate is effortless with several flexible set-up options to choose from. Once an assay is run, data can be transferred in Excel-compatible .csv format from the Modulus™ II Microplate using the built-in USB drive and the USB stick that comes with the instrument. Affordable Modularity The Modulus™ II Microplate is a modular instrument that easily fits into most budgets. Purchase the technology or modes that you need now and add onto the system later as your needs expand. For example, the Modulus™ II Microplate can be purchased as a luminometer, then fluorescence and/or absorbance modules can be purchased and added later. There is no need for a service call or downtime. Installations take less than ten minutes and can be done right in your lab using just an Allen wrench which is provided with the module. In addition, after installing a new module, you will not need to download and install new software. The Modulus™ II Microplate instantly detects the newly installed module(s) and will automatically adjust screens, protocols, and options.
Figure 2: The Fluorescence Module uses an epifluorescent design to ensure a consistent measurement position. This approach reduces the impact of small sample volume fluctuations by consistently measuring the sample at the same position from well to well.
Figure 3: The Modulus™ II Multimode Microplate Reader provides scientists with two different options for photometric measurements. The UV-Visible Absorbance Module allows measurements from 200 - 1100 nm. The second option is the Visible Absorbance Module allowing measurements from 400 - 800 nm. Microplate Formats The Modulus™ II Multimode Microplate Reader accepts 6, 12, 24, 48, 96, and 384-well plates conforming to the SBS plate standard.
Data Handling The Modulus™ II Microplate provides a simple and unique method for data handling. A USB stick is included and a USB drive is built into the instrument. To transfer data to either PC or Mac, simply copy the data from the Modulus™ II Microplate then move the USB stick to your computer. This method allows you to analyze your data when and where you find it most convenient. Touch Screen The Modulus™ II Microplate combines a 6.6” color touch screen display with an onboard Windows-based computer. In addition, the intuitive user interface makes setting up a run and retrieving data fast and simple while maintaining the flexibility needed for advanced or custom protocols.
Temperature Control The Modulus™ II Multimode Microplate Reader may be purchased with an optional heater allowing for precise temperature control from 2º above ambient temperature to 45º C. Shaking The Modulus™ II Multimode Microplate Reader is configured with a factory-installed shaker that allows for either linear or orbital shaking. Protocol Composer The Modulus™ II Multimode Microplate Reader comes with a Protocol Composer. The Protocol Composer allows complex protocols to be easily developed by combining multiple technologies into one experiment. Save time and reagents by testing for both luminescence and fluorescence or fluorescence and absorbance in one run. Luminescence Channel Applications Features • > 8 logs dynamic range
*Recommended Luminescence Module(Factory Installed) The Modulus™ II Multimode Microplate Reader with the Luminescence Module is designed to deliver performance equivalent to dedicated microplate luminometers while also offering the flexibility of a multimode reader. To achieve sensitivity on par with that of a dedicated luminometer, the luminescence channel is separated from other measurement technologies and positioned directly above the sample well. These conditions maximize light capture for the best possible sensitivity. In conjunction, a low-noise photomultiplier tube ensures that collected light is not compromised in any way. This design makes the Modulus™ II Microplate between 10 to 1000 times more sensitive than competing multimode luminometers. In addition to high sensitivity, the Modulus™ II Microplate boasts greater than 8 logs of dynamic range, eliminating the need to dilute samples or manage detector-driven gain changes. To achieve this extra large reading range, the Modulus™ II Microplate is capable of simultaneously measuring samples of varying brightness. The photomultiplier tube automatically adjusts for the optimum reading of bright or dim samples. This means that the Modulus™ II Microplate is capable of achieving a reading range of 2 – 3 logs more than competing multimode luminometers. For further functionality, a Modulus™ II Microplate with the Luminescence Module is designed to reduce crosstalk. The luminescent light path is isolated through a dual-masking system at both the detector and the well being read. Dual masking eliminates stray light by creating a column in which the light from the sample passes directly to the detector. When using 96-well white plates, this dual-masking system effectively results in a crosstalk rejection of 5 x 10-5. Fluorescence Channel Applications Features • Epifluorescent detection
Fluorescence Module (User Installable) The Modulus™ II Multimode Microplate Reader with the Fluorescence Module installed is designed to deliver both high performance and user flexibility. To achieve high performance, the Fluorescence Module utilizes powerful light-emitting diodes (LEDs) as excitation sources. LEDs have very specific light-output profiles which closely match the excitation profiles of commonly used fluorescent molecules. LED usage increases sensitivity by fully exciting the fluorophore and reducing non-specific light leakage, a problem often found when using broad-spectrum light sources. Additionally, the Modulus™ II Microplate Fluorescence Module uses an epifluorescent design to ensure a consistent measurement position. This pinpointed approach reduces the impact of volume fluctuation within small samples by reliably measuring at the same position from well to well. The Modulus™ II Microplate Fluorescence Module accepts different Fluorescence Optical Kits for measuring a variety of fluorophores. Four standard optical kits are available to measure the most popular fluorophores including Hoechst dye, fluorescein, rhodamine, Cy3, and Cy5. In addition, custom optical kits can be readily made for non-standard applications. Optical kits can be easily exchanged in seconds and built-in software ensures that the installed optical kit matches the selected application protocol. Protocols for nucleic acids and proteins such as PicoGreen®, RiboGreen®, and Quant-iT™ assays are preprogrammed into the Modulus™ II Microplate for your convenience. Absorbance Channel Applications Features • UV-Visible or Visible Modules
Absorbance Module (User Installable) The Modulus™ II Multimode Microplate Reader provides scientists with two different options for making photometric measurements. Both modules are user installable, so there is no downtime or need for a service call. The Modulus™ II Microplate with an Absorbance Module provides measurements that are highly sensitive and cover a wide dynamic range. In addition, it is easy to run ratiometric absorbance-based assays to adjust for optical imperfections or to assess DNA purity. Simply use the color touch screen to select two desired wavelengths. The instrument has an easy-to-read screen that displays Optical Density (OD) values for each wavelength or a ratio of two wavelengths.
Injectors
Luminescence Light Plate
Data Analysis Because data is provided in .csv format, any necessary data manipulations can be done within Excel. For optimum ease of analysis, Turner BioSystems has additionally developed optional, specialized software. This Curve-Fitting Data Analysis Software program provides calculation, graphing, and printing of eight different curve-fitting methods: linear fit, quadratic fit, cubic fit, two-parameter fit, four-parameter with linear x-axis fit, four-parameter with log two-axis fit, linear spline, and cubic spline. This software option is compatible with the Windows XP operating system for PC computers. Computer Interface An optional software program is available for those labs that would prefer to operate the Modulus™ II Multimode Microplate Reader through an external PC. The PC Connect Kit contains all of the same ease-of-use features which are available via the instrument’s built-in touch screen. Specifications Modulus™ II Multimode Microplate Reader Instrument Specifications (9310-010) Detection Modes: Luminescence, Fluorescence, Absorbance Modulus™ II Multimode Microplate Reader Instrument with Heating (9310-011) Temperature Control: 2 ºC above ambient temperature to 45 ºC +/- 0.75 ºC Luminescence Module Specifications (9310-020 Factory Installed) Detector: Head-on photomultiplier tube (PMT) for photon counting Fluorescence Module Specifications (9310-040 User Installable) Light Source: Wavelength-matched LED Visible Absorbance Module Specifications (9310-050 User Installable) Light Source: LED UV-Visible Absorbance ModuleSpecifications (9310-051 User Installable) Light Source: Xenon lamp Single Injector System Specifications (9300-061 Optional) Number of Injectors: One injector Dual Injector System Specifications (9300-062 Optional) Number of Injectors: Two injectors Terms of iPod Promotion:
|
| ||||||||||||||||||||||||||||||||||||||||||||||||||
| Promega Corporation Sunnyvale Office • Detection Instruments 645 N Mary Avenue • Sunnyvale CA 94085 Phone: 408.636.2400 - Toll Free: 888.636.2401 - Fax: 408.737.7919 |
||
|
||